Missing haemodynamic measurement codes
Of the haemodynamic measurements described on page 167 of the enclosed manual LAB-11384A_Master.pdf we're unable to identify nomenclature codes for the following:
- SVV: Stroke Volume Variation (The maximum stroke volume minus the minimum stroke volume divided by the mean stroke volume across at least one respiratory cycle (10 seconds))
- PPV: Pulse Pressure Variation (The maximum PP minus the minimum PP divided by the mean PP across at least one respiratory cycle (10 seconds))
- DO2: Oxygen Delivery Index (The product of cardiac output and oxygen concentration. The DO2 is the amount of oxygen delivered to the tissues.)
- HRV: Heart Rate Variation (The standard deviation in heart rate over a respiratory cycle (10 sec) divided by the mean HR)
Would it be possible to include these in the upcoming 10101b?